FDA Adverse Event Malfunction Summary report: N

BRACHY STAR SEED IMPLANT NEEDLE

MDR report key: 4349338 · Received December 12, 2014

Report

Report Number
1018233-2014-00352
Event Type
Malfunction
Date Received
December 12, 2014
Report Date
October 27, 2016
Manufacturer
BARD REYNOSA S.A. DE C.V.
Product Code
IWJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEEDLE BROKE WHILE IN THE PT. THE NEEDLE CONTAINED 6 SEEDS WHEN IT BROKE DURING THE IMPLANT. THE NEEDLE WAS IN THE PT WHEN IT BROKE BUT THEY WERE ABLE TO RETRIEVE BOTH PARTS OF THE NEEDLE AND THE SEEDS WERE STILL IN THE NEEDLE. PER ADDITIONAL INFO RECEIVED THE NEEDLE WAS BEING MANIPULATED BETWEEN THE PT AND GRID WITH A DIDDLER" (TOOL USED TO HELP GUIDE THE NEEDLE TIP TO THE TARGET AREA OF THE PROSTATE). THE NEEDLE WAS IN THE PT'S PERINIUM APPROXIMATELY 1-2 CM WHEN THE BREAKAGE OCCURRED LEAVING 2-3 CM PROTRUDING OUTSIDE OF THE PT'S BODY. THE PHYSICIAN WAS ABLE TO GRASP THE NEEDLE WITH FORCEPS AND PULL OUT THE BROKEN PORTION FROM THE PT. THE DOCTOR PERFORMED A FLUOROSCOPY TO ENSURE THAT NO FRAGMENTS REMAINED IN THE PT. NO FRAGMENTS WERE NOTED.