FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4345488 · Received December 5, 2014

Report

Report Number
1720753-2014-09598
Event Type
Malfunction
Date Received
December 5, 2014
Date of Event
November 17, 2014
Report Date
December 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE GENERATOR INTERFACE PCB (GIB) AND FLUOROSCOPY FUNCTIONS PCB (FFB) WERE EVALUATED AND RESEATED. CONNECTIONS INSIDE THE PWB RACK BOX WERE ALSO RESET DURING THE SERVICE EVENT. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PATIENT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786289 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1