FDA Adverse Event Malfunction Summary report: N

COBAS 8000 E602 MODULE

MDR report key: 4330829 · Received December 15, 2014

Report

Report Number
1823260-2014-09985
Event Type
Malfunction
Date Received
December 15, 2014
Date of Event
December 5, 2014
Report Date
January 5, 2015
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K100853
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SPECIFIC ROOT CAUSE COULD NOT BE IDENTIFIED. ADDITIONAL INFORMATION FOR FURTHER INVESTIGATION WAS REQUESTED BUT WAS NOT PROVIDED. A GENERAL INSTRUMENT OR REAGENT ISSUE WAS UNLIKELY AS NO ISSUES WERE FOUND BY THE FIELD SERVICE REPRESENTATIVE, THE ALARM TRACE INDICATED NO ISSUES, AND CALIBRATION AND QC WERE ACCEPTABLE.

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER RECEIVED A QUESTIONABLE TESTOSTERONE RESULT FOR ONE PATIENT SAMPLE. THE INITIAL RESULT WAS 12.5 NG/DL AND WAS REPORTED OUTSIDE THE LABORATORY. THE DOCTOR QUESTIONED THE RESULT AS THE RESULT FOR THIS PATIENT LAST MONTH WAS 454 NG/DL. THE SAME POUR OFF TUBE WAS REPEATED AND THE RESULT WAS 427 NG/DL ON ANOTHER ANALYZER. THEY REPEATED THE SAMPLE MULTIPLE TIMES AND GOT RESULTS BETWEEN 400 AND 450 NG/DL ON VARIOUS MODULES. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT. THERE WAS NO ADVERSE EFFECT TO PATIENT. THE REAGENT LOT NUMBER WAS 17878202 WITH AN EXPIRATION DATE OF 10/31/2015. THE FIELD SERVICE REPRESENTATIVE COULD NOT FIND A PROBLEM. HE EXAMINED THE UNIT AND RAN PERFORMANCE TESTING WITH RESULTS WITHIN SPECIFICATION. HE CHECKED AND INSPECTED THE INSTRUMENT AND CHECKED THE FLUID AND WASH CYCLES. HE REVIEWED THE CALIBRATION AND CONTROLS WHICH WERE OK. HE CHECKED THE SAMPLE INTEGRITY WHICH WAS OK. HE CHECKED THE SAMPLE PROBE AND LLD VOLTAGE AND FOUND THE SAMPLE TIP WAS SLIGHTLY DIRTY. HE CLEANED IT AND RECHECKED THE VOLTAGE WHICH WAS OK. HE REVIEWED THE CLEANING PROCEDURE FOR THE SAMPLE PROBE TIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
819853 COBAS 8000 E602 MODULE IMMUNOCHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1 044 YR