FDA Adverse Event Injury Summary report: N

MILLENNIUM VG

MDR report key: 4329331 · Received December 15, 2014

Report

Report Number
9613299-2014-00005
Event Type
Injury
Date Received
December 15, 2014
Date of Event
November 19, 2014
Report Date
November 19, 2014
Manufacturer
GE MEDICAL SYSTEMS ISRAEL FUNCTIONAL IMAGING
Product Code
KPS
PMA / PMN Number
K953801
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
SERVICE PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION REVEALED THERE WAS NO DEFECT OF THE SYSTEM INVOLVED OR ITS PLANNED MAINTENANCE SERVICE INSTRUCTIONS. THE SERVICE ENGINEER FAILED TO PROPERLY FOLLOW EHS GUIDELINES, DID NOT OBSERVE THE AREA BEFORE INSERTING HIS HAND AND ALSO DID NOT USE PROPER EQUIPMENT (CUT RESISTANT GLOVES). SERVICE PERSONNEL ARE TRAINED ON NUC SYSTEMS AND EHS GUIDELINES. THESE GUIDELINES INCLUDE USE OF PPE IN THE FORM OF CUT RESISTANT GLOVES FOR FRU REPLACEMENT THAT INVOLVES REPLACEMENT OF PARTS IN TIGHT SPACES THAT HAVE THE POTENTIAL OF PINCHING. THE SERVICE ENGINEER WAS WEARING SURGICAL GLOVES BUT NO CUT RESISTANT GLOVES.

Additional Manufacturer Narrative · 1

DATE OF MANUFACTURE UNKNOWN. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

A FIELD ENGINEER (FE) WAS PERFORMING PLANNED MAINTENANCE ON A VG SYSTEM AT A CUSTOMER LOCATION. THE FE WAS TIGHTENING A SCREW USING AN ALLEN WRENCH, THE ALLEN WRENCH SLIPPED OFF CAUSING HIS HAND TO COME IN CONTACT WITH A SHARP EDGE. AS A RESULT, THE FE SUSTAINED A CUT ON ONE OF HIS FINGERS, REQUIRING THREE STITCHES AND A TETANUS SHOT. THE FE WAS WEARING SURGICAL GLOVES BUT NO CUT RESISTANT GLOVES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
821497 MILLENNIUM VG TOMOGRAPHY COMPUTED EMISSION SYSTEM KPS GE MEDICAL SYSTEMS ISRAEL FUNCTIONAL IMAGING

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other