FDA Adverse Event Malfunction Summary report: N

NOVA MAX TEST STRIPS

MDR report key: 4323190 · Received December 5, 2014

Report

Report Number
MW5039465
Event Type
Malfunction
Date Received
December 5, 2014
Date of Event
September 9, 2014
Report Date
December 5, 2014
Manufacturer
NOVA BIOMEDICAL
Product Code
NBW
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
PHARMACIST

Narratives

Description of Event or Problem · 1

PATIENT STATES HAS BEEN RECHECKING BG WITHIN 1 MINUTE TO DOUBLE CHECK RESULT AND NOTICED THAT THE VALUE IS DIFFERENT (84 --> 193, 152 -->84, 212-->163, 193-->160). PT ALSO EXPRESSES CONCERN THAT GLUCOMETER IS NOT WORKING PROPERLY. IN CLINIC, USE CONTROL SOLUTION AND APPLIED TO PATIENT'S TEST STRIP. THE RESULT WAS 171 WHICH WAS NOT WITHIN GIVEN RANGE BY THE MANUFACTURER OF 82-127 MG/DL. ALSO TESTED THE PATIENT'S METER USING CLINIC TEST STRIP AND THE RESULT WAS WITHIN GIVEN RANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
786558 NOVA MAX TEST STRIPS NOVA MAX TEST STRIPS NBW NOVA BIOMEDICAL 1020214027

Patients

Seq Age Sex Outcome Treatment
1 52 YR