FDA Adverse Event
Malfunction
Summary report: N
NOVA MAX TEST STRIPS
MDR report key: 4323190
·
Received December 5, 2014
Report
- Report Number
- MW5039465
- Event Type
- Malfunction
- Date Received
- December 5, 2014
- Date of Event
- September 9, 2014
- Report Date
- December 5, 2014
- Manufacturer
- NOVA BIOMEDICAL
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- PHARMACIST
Narratives
Description of Event or Problem · 1
PATIENT STATES HAS BEEN RECHECKING BG WITHIN 1 MINUTE TO DOUBLE CHECK RESULT AND NOTICED THAT THE VALUE IS DIFFERENT (84 --> 193, 152 -->84, 212-->163, 193-->160). PT ALSO EXPRESSES CONCERN THAT GLUCOMETER IS NOT WORKING PROPERLY. IN CLINIC, USE CONTROL SOLUTION AND APPLIED TO PATIENT'S TEST STRIP. THE RESULT WAS 171 WHICH WAS NOT WITHIN GIVEN RANGE BY THE MANUFACTURER OF 82-127 MG/DL. ALSO TESTED THE PATIENT'S METER USING CLINIC TEST STRIP AND THE RESULT WAS WITHIN GIVEN RANGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 786558 | NOVA MAX TEST STRIPS | NOVA MAX TEST STRIPS | NBW | NOVA BIOMEDICAL | 1020214027 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |