FDA Adverse Event Malfunction Summary report: N

TRIDENT HEMISPHERICAL SOLID BK 50MM

MDR report key: 4322373 · Received December 11, 2014

Report

Report Number
0002249697-2014-04589
Event Type
Malfunction
Date Received
December 11, 2014
Date of Event
November 12, 2014
Report Date
November 12, 2014
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K013676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVENT REGARDING A PACKAGING ISSUE INVOLVING A TRIDENT SHELL WAS REPORTED. THE EVENT WAS NOT CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: NOT PERFORMED, NO PACKAGING OR DEVICES WERE RETURNED. MEDICAL RECORDS RECEIVED AND EVALUATION: INSUFFICIENT MEDICAL RECORDS WERE RECEIVED FOR REVIEW WITH A CLINICAL CONSULTANT. DEVICE HISTORY REVIEW: ALL DEVICES IN THE REPORTED LOT WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS FOR THE REPORTED LOT. CONCLUSIONS: THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE INSUFFICIENT INFORMATION WAS PROVIDED. FURTHER INFORMATION SUCH AS RETURN OF THE REPORTED PACKAGING IS NEEDED TO COMPLETE THE INVESTIGATION FOR DETERMINING ROOT CAUSE.

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. PRODUCT DISPOSED AT HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OUTER STERILE PACKAGING WAS STUCK TO INNER STERILE PACKAGING. SURGEON WAS NOT HAPPY USING IT. PER ADDITIONAL INFORMATION FROM THE REP: NO ADVERSE EVENT AS ANOTHER PRODUCT WAS AVAILABLE FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OUTER STERILE PACKAGING WAS STUCK TO INNER STERILE PACKAGING. SURGEON WAS NOT HAPPY USING IT. PER ADDITIONAL INFORMATION FROM THE REP: NO ADVERSE EVENT AS ANOTHER PRODUCT WAS AVAILABLE FOR USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
805017 TRIDENT HEMISPHERICAL SOLID BK 50MM IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH 47190101

Patients

Seq Age Sex Outcome Treatment
1 Other