FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4321237 · Received December 2, 2014

Report

Report Number
3008642652-2014-04642
Event Type
Malfunction
Date Received
December 2, 2014
Date of Event
November 4, 2014
Report Date
December 1, 2014
Manufacturer
ZOLL MANUFACTURING CORP
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. UPON EVALUATION, THE BATTERY CHARGER/MODEM WOULD NOT POWER ON. THE CAUSE OF THE PROBLEM WAS ISOLATED TO INTERMITTENT U104 (PXA) AND U1003 (PLD) COMPONENTS ON COMPUTER ANALOG (CA) BOARD SN (B)(4). THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A ZOLL DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) AND REPORTED THAT THE CHARGER/MODEM WAS RESETTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
776351 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTE RDEFIBRILLATOR MVK ZOLL MANUFACTURING CORP WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA