LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2014-04642
- Event Type
- Malfunction
- Date Received
- December 2, 2014
- Date of Event
- November 4, 2014
- Report Date
- December 1, 2014
- Manufacturer
- ZOLL MANUFACTURING CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. UPON EVALUATION, THE BATTERY CHARGER/MODEM WOULD NOT POWER ON. THE CAUSE OF THE PROBLEM WAS ISOLATED TO INTERMITTENT U104 (PXA) AND U1003 (PLD) COMPONENTS ON COMPUTER ANALOG (CA) BOARD SN (B)(4). THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.
A ZOLL DISTRIBUTOR RETURNED BATTERY CHARGER/MODEM SN (B)(4) AND REPORTED THAT THE CHARGER/MODEM WAS RESETTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 776351 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTE RDEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORP | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |