FDA Adverse Event Injury Summary report: N

CLINITRON AT HOME

MDR report key: 4320444 · Received November 26, 2014

Report

Report Number
1824206-2014-02720
Event Type
Injury
Date Received
November 26, 2014
Date of Event
October 31, 2014
Report Date
October 31, 2014
Manufacturer
HILL-ROM, INC.
Product Code
INX
PMA / PMN Number
K942184
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BLOWER WAS INOPERATIVE. THE TECHNICIAN REPLACED THE BLOWER AND THE ISSUE WAS RESOLVED. THE REPORTED INJURY IS SERIOUS IN NATURE PER FDA DEFINITION. HOWEVER, THE DETERIORATION OF THE PRESSURE ULCER TOOK PLACE WHILE THE PATIENT WAS NOT ON A HILL-ROM SURFACE. DEVELOPMENT OF PRESSURE ULCERS IS MULTIFACTORIAL AND CANNOT BE ONLY ATTRIBUTED TO PERFORMANCE OF THE SURFACE. RISK FACTORS INCLUDE PROTEIN-CALORIE MALNUTRITION, MICROCLIMATE (SKIN WETNESS CAUSED BY SWEATING OR INCONTINENCE), DISEASES THAT REDUCE BLOOD FLOW TO THE SKIN, SUCH AS ARTERIOSCLEROSIS, OR DISEASES THAT REDUCE THE SENSATION IN THE SKIN, SUCH AS PARALYSIS OR NEUROPATHY. POSITION CHANGES ARE KEY TO PRESSURE SORE PREVENTION AND TREATMENT. THESE CHANGES NEED TO BE FREQUENT, REPOSITIONING NEEDS TO AVOID STRESS ON THE SKIN, AND BODY POSITIONS NEED TO MINIMIZE THE RISK OF PRESSURE ON VULNERABLE AREAS.

Description of Event or Problem · 1

THE PATIENT'S WIFE REPORTED THAT HER HUSBAND'S WOUNDS WERE GETTING WORSE AND HE DEVELOPED A NEW WOUND. THE PATIENT IS A DOUBLE AMPUTEE AT THE PELVIS. THE ACCOUNT STATED THAT THE BED ALARMED AND SHUT DOWN AND THEN THEY CONTACTED HILL-ROM. THE TECHNICIAN EVALUATED THE UNIT AND FOUND THE PART NEEDED WAS NOT IN STOCK AND LET THE ACCOUNT KNOW. THE ACCOUNT THEN PLACED THE PATIENT ON A NON-HILL-ROM MATTRESS FOR 7 DAYS, UNTIL THE REPAIR COULD BE COMPLETED. THE ACCOUNT TREATED THE WOUNDS WITH SILVADENE AND TWO TYPES OF ANTIBIOTICS. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
769196 CLINITRON AT HOME BED, AIR-FLUIDIZED INX HILL-ROM, INC. 0800

Patients

Seq Age Sex Outcome Treatment
1 59 YR