EASYPUMP: LT 270-132: 270 ML, 22 ML/HR
Report
- Report Number
- 2026095-2014-00226
- Event Type
- Malfunction
- Date Received
- November 7, 2014
- Date of Event
- July 7, 2014
- Report Date
- October 8, 2014
- Manufacturer
- I-FLOW LLC
- Product Code
- MEB
- PMA / PMN Number
- K052117
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IT WAS REPORTED THAT ALL 5 SUSPECT DEVICES WERE DISCARDED; THEREFORE, THE REPORTER RETURNED 5 REPRESENTATIVE SAMPLES FROM THE SAME LOT NUMBER FOR TESTING. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED FOR THE LOT NUMBER REPORTED. EVAL AND INVESTIGATION RESULTS WILL BE PROVIDED ONCE THEY ARE COMPLETED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED FOR THE LOT NUMBER OF THE MANUFACTURED UNIT. THERE WERE NO REWORKS, SPECIAL CONDITIONS, OR RELATED NONCONFORMITY REPORTS (NCRS) FOR THIS LOT. THE LOT MET THE PROCESS SPECIFICATIONS, INCLUDING THE QUALITY CONTROL ACCEPTANCE CRITERIA PRIOR TO RELEASE. A CONCLUSION IS NOT YET AVAILABLE AS THE INVESTIGATION AND EVALUATION OF THE DEVICES IS STILL IN PROGRESS. ONCE THE INVESTIGATION AND DEVICE ANALYSIS ARE COMPLETED A FOLLOW UP REPORT WILL BE SUBMITTED. INFO FROM THIS INCIDENT HAS BEEN INCLUDED IN OUR PRODUCT COMPLAINT AND MDR TREND REPORTING SYSTEMS. TREND INFO IS USED TO IDENTIFY THE NEED FOR ADDITIONAL INVESTIGATIONS.
FILL VOLUME 330 ML. FLOW RATE 2 ML/HR. PROCEDURE: CHEMOTHERAPY. CATHPLACE: RIGHT CHEST (PORT-A-CATH). REPORT 4 OF 5. PLEASE REFER TO 2026095-2014-00214/14-00885 (A), 2026095-2014-00224/14-00885 (B), 2026095-2014-00225/14-00885 (C), 2026095-2014-00227/14-00885 (E). PUMP 4. AN INCIDENT INVOLVING ONE PT WITH 5 INCIDENTS OF FAST FLOWS WERE REPORTED. THE INCIDENT WAS DESCRIBED AS "INFUSER FINISHED AHEAD OF SCHEDULE (IN THE SAME PT)." IT WAS REPORTED THAT THE INFUSION ENDED 1 OR 2 DAYS EARLIER THAN EXPECTED. THE PT REPORTED FATIGUE, NAUSEA AND MALAISE FOLLOWING THE INFUSIONS. AT THE TIME OF THIS REPORT, THE PT WAS STILL UNDERGOING TREATMENT. THE PUMP WAS FILLED ON (B)(6) 2014 AND INFUSION WAS STARTED ON THE SAME DAY. THE INFUSION ENDED ON (B)(6) 2014 (5 DAYS), THIS IS WHEN THE DEFECT/INCIDENT WAS NOTICED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 720881 | EASYPUMP: LT 270-132: 270 ML, 22 ML/HR | MEB | I-FLOW LLC | 04434376 | 0201022114 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | DRUG: FLUORACILA 8.92 ML DILUTED IN NORMAL SALINE |