FDA Adverse Event Injury Summary report: N

BO-HLS 7050 #HIT SET ADVANCED 7.0

MDR report key: 4317506 · Received December 2, 2014

Report

Report Number
8010762-2014-01322
Event Type
Injury
Date Received
December 2, 2014
Date of Event
November 3, 2014
Report Date
November 3, 2014
Manufacturer
MAQUET CARDIOPULMONARY AG
Product Code
DTZ
PMA / PMN Number
K101153
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MAQUET CARDIOPULMONARY IS AWARE OF SIMILAR COMPLAINTS. THE DEVICES DISPLAYED A SIMILAR MALFUNCTION WHICH WERE TESTED AND EVALUATED UNDER AN OPTICAL MICROSCOPE. DELAMINATION OF SOME GAS FIBERS WAS OBSERVED WHICH ALLOWED FOR THE PRIMING SOLUTION OR BLOOD TO FLOW INSIDE THE GAP BETWEEN THE GAS FIBERS AND POLYURETHANE. GRAVITY THEN ALLOWED FOR PASSAGE TO THE GAS EXITING PATH ALONG THE HOUSING. THE MOST PROBABLE ROOT-CAUSE IS THE DELAMINATION OF THE HOLLOW GAS FIBERS FROM THE POLYURETHANE POTTING AREA. A REVIEW OF THE QC PROCESS CONFIRMS THAT 100% FUNCTIONAL INSPECTION FOR LEAKAGE IS PERFORMED DURING PRODUCTION. MAQUET CARDIOPULMONARY (B)(4) HAS INITIATED AN INTERNAL PROCESS ((B)(4)) TO ADDRESS THE APPROPRIATE CORRECTIVE AND PREVENTIVE ACTION. ADDITIONAL INFO: THE PRODUCT MENTIONED IS NOT DISTRIBUTED TO THE UNITED STATES, BUT THE PRODUCT WITH CONTRIBUTING DESIGN FUNCTION TO THE AFFECTED COMPONENT (QUADROX-ID) IS REGISTERED UNDER 510(K): K101153. A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED WHEN NEW INFO BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT SHORTLY AFTER INITIATING ECMO, LEAKAGE FROM THE GAS OUTLET WAS NOTED. THE DEVICE WAS EXCHANGED. NO PT EFFECT WAS REPORTED. ECMO - EXTRA CORPOREAL MEMBRANE OXYGENATION. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
776067 BO-HLS 7050 #HIT SET ADVANCED 7.0 TUBING PACK DTZ MAQUET CARDIOPULMONARY AG 70104.9134

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention