FDA Adverse Event Death Summary report: N

M1176A COMPONENT MONITORING SYSTEM

MDR report key: 431732 · Received December 10, 2002

Report

Report Number
9610816-2002-00189
Event Type
Death
Date Received
December 10, 2002
Date of Event
December 11, 2000
Report Date
October 9, 2002
Manufacturer
AGILENT TECHNOLOGIES DEUTSCHLAND GMBH
Product Code
DRT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

IT WAS REPORTED TO PHILIPS MEDICAL SYSTEMS THAT THE ANESTHESIA MACHINE, WHICH PROVIDED OXYGEN TO THE PT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, WAS SHUT OFF IN ORDER TO TAKE A CHOLANGIOGRAM. IT IS ALLEGED THAT THE ALARM SYSTEM ON THE PT MONITOR WAS PROGRAMMED IN THE "ALARMS SUSPENDED" MODE, WHICH TURNED OFF ALL AUDIBLE MONITOR ALARMS DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M1176A COMPONENT MONITORING SYSTEM BEDSIDE MONITOR DRT AGILENT TECHNOLOGIES DEUTSCHLAND GMBH M1094B NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Death