FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 4314424 · Received December 10, 2014

Report

Report Number
2649622-2014-15279
Event Type
Injury
Date Received
December 10, 2014
Date of Event
October 21, 2014
Report Date
October 21, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THIS DATA INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD. IT WAS NOTED THAT THERE WERE 505 V-SIC (VENTRICULAR- SHORT INTERVAL COUNTS) THAT OCCURRED SINCE 2014-(B)(6) AND NINE NSVT (NON-SUSTAINED VENTRICULAR TACHYCARDIA) EPISODES OF LESS THAN 220 MN V-V CYCLE WERE RECORDED ON FIVE DAYS DURING (B)(6) 2014.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. IT WAS FURTHER REPORTED THAT ANALYSIS OF THIS DATA INDICATED THAT THERE OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER) ON THE RIGHT VENTRICULAR (RV) LEAD. THERE WERE 10 NST (NON-SUSTAINED TACHYCARDIA) EPISODES WITH V-V INTERVALS LESS THAN 220 MS FROM (B)(4) 2015. ADDITIONALLY, THERE WAS 4017 V-SICS SINCE LAST SESSION 189 DAYS AGO. IT WAS NOTED PACING INHIBITED DUE TO OVERSENSING.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: A2DR01 IPG IMPLANTED: (B)(6) 2014. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD OVERSENSING AS IT SHOWED TO HAVE FIVE HUNDRED AND TWO SHORT V-V¿S. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD CONTINUED TO EXHIBIT OVERSENSING V-SIC (VENTRICULAR - SENSING INTEGRITY COUNTER) IS GREATER THAN FOUR THOUSAND. ADDITIONALLY, IT WAS NOTED THAT THE EPISODES WITH EGM (ELECTROGRAM) AVAILABLE APPEAR TO BE OVERSENSING WHICH SHOW INHIBITION OF PACING. THEY WERE UNABLE TO REPRODUCE OVERSENSING WITH ISOMETRICS OR POCKET MANIPULATION. THE SENSITIVITY WAS PROGRAMMED AND THE LEAD WAS ALSO REPROGRAMMED TO UNIPOLAR. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
798090 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00069 YR Hospitalization| R