FDA Adverse Event Malfunction Summary report: N

DA VINCI S SURGICAL SYSTEM

MDR report key: 4309934 · Received December 9, 2014

Report

Report Number
2955842-2014-05821
Event Type
Malfunction
Date Received
December 9, 2014
Date of Event
November 4, 2014
Report Date
November 17, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL INC. (ISI) RECEIVED THE PSM INVOLVED WITH THIS COMPLAINT AND COMPLETED THE INVESTIGATION. FAILURE ANALYSIS FOUND THE PSM HAD FAILED THE WEIGHTED BRAKE DROP TEST.

Description of Event or Problem · 1

DURING A PREVENTATIVE MAINTENANCE OF A DA VINCI SI SURGICAL SYSTEM BY AN INTUITIVE SURGICAL INC. REPRESENTATIVE FOUND THE PATIENT SIDE MANIPULATOR (PSM) HAD FAILED THE BRAKE TEST. THE PSM WAS REPLACED. THE PSM IS AN INSTRUMENT ARM WHICH IS LOCATED ON THE PATIENT SIDE CART THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
795917 DA VINCI S SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS2000 A5.1P8

Patients

Seq Age Sex Outcome Treatment
1