DA VINCI S SURGICAL SYSTEM
MDR report key: 4309934
·
Received December 9, 2014
Report
- Report Number
- 2955842-2014-05821
- Event Type
- Malfunction
- Date Received
- December 9, 2014
- Date of Event
- November 4, 2014
- Report Date
- November 17, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INTUITIVE SURGICAL INC. (ISI) RECEIVED THE PSM INVOLVED WITH THIS COMPLAINT AND COMPLETED THE INVESTIGATION. FAILURE ANALYSIS FOUND THE PSM HAD FAILED THE WEIGHTED BRAKE DROP TEST.
Description of Event or Problem · 1
DURING A PREVENTATIVE MAINTENANCE OF A DA VINCI SI SURGICAL SYSTEM BY AN INTUITIVE SURGICAL INC. REPRESENTATIVE FOUND THE PATIENT SIDE MANIPULATOR (PSM) HAD FAILED THE BRAKE TEST. THE PSM WAS REPLACED. THE PSM IS AN INSTRUMENT ARM WHICH IS LOCATED ON THE PATIENT SIDE CART THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 795917 | DA VINCI S SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS2000 A5.1P8 |