FDA Adverse Event Malfunction Summary report: N

DA VINCI SI

MDR report key: 4309913 · Received December 9, 2014

Report

Report Number
2955842-2014-05820
Event Type
Malfunction
Date Received
December 9, 2014
Date of Event
October 31, 2014
Report Date
November 10, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K081137
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL INC. (ISI) RECEIVED THE PSM INVOLVED WITH THIS COMPLAINT AND COMPLETED THE INVESTIGATION. FAILURE ANALYSIS CONFIRMED THE REPORTED ISSUE OF THE PSM FAILING FRICTION TEST. THE PSM ALSO FAILED THE WEIGHTED BRAKE DROP TEST.

Description of Event or Problem · 1

DURING A PREVENTATIVE MAINTENANCE OF A DA VINCI SI SURGICAL SYSTEM BY AN INTUITIVE SURGICAL INC. REPRESENTATIVE FOUND THE PATIENT SIDE MANIPULATOR (PSM) HAD FAILED THE FRICTION TEST ON THE PSM. THE PSM WAS REPLACED. THE PSM IS AN INSTRUMENT ARM WHICH IS LOCATED ON THE PATIENT SIDE CART THAT PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
793951 DA VINCI SI ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL,INC. IS3000 A6.0P8