FDA Adverse Event
Summary report: N
ACCUTRON
MDR report key: 430489
·
Received December 2, 2002
Report
- Report Number
- MW1026863
- Date Received
- December 2, 2002
- Date of Event
- October 1, 2002
- Report Date
- November 19, 2002
- Manufacturer
- ACCUTRON
- Product Code
- KHA
- Report Source
- Voluntary report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
Narratives
Description of Event or Problem · 1
AN INCIDENT OCCURRED IN THE OFFICE WITH A N2O-O2 UNIT THAT WAS RECENTLY INSTALLED. THE PIPES WERE CROSSED IN THE WALLS. DENTIST WAS GIVING PATIENTS 100% N2O WHEN THEY THOUGHT THEY WERE GIVING THEM 100% O2. THE PROBLEM WAS COMPOUNDED BY THE USE OF A FLAVORED MASK. THE SCENT INTERFERED WITH THE CHARACTERISTIC ODOR OF N2O. DENTIST COULD NOT SMELL THE ODOR OF N2O. THEY COULD ONLY SMELL THE SCENT FROM THE FLAVORED MASK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCUTRON | NASAL HOOD WITH SCENTS | KHA | ACCUTRON | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |