IMMULITE 2000 ANDROSTENEDIONE
Report
- Report Number
- 2432235-2014-00687
- Event Type
- Malfunction
- Date Received
- December 5, 2014
- Date of Event
- July 2, 2014
- Report Date
- November 14, 2014
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS LIMITED
- Product Code
- CIZ
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- 2432235-11/20/14-005R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SIEMENS HAS OBSERVED OVER-RECOVERY IN THE IMMULITE/IMMULITE 1000/IMMULITE 2000 ANDROSTENEDIONE ASSAY FOR SAMPLES ACROSS THE ASSAYS' REPORTABLE RANGE OF 0.3-10 NG/ML (1.1-35 NMOL/L). SIEMENS HAS CONFIRMED THAT THE ISSUE NOTED IN THE DECEMBER 2013 FIELD ACTION IS NOT LIMITED TO ONLY SAMPLES >5.5 NG/ML (19.2 NMOL/L). SIEMENS HAS IDENTIFIED THAT THE ROOT CAUSE OF THE OVER-RECOVERY IS RELATED TO THE VARIABILITY OF A CRITICAL RAW MATERIAL. THIS ISSUE HAS BEEN RESOLVED BEGINNING WITH IMMULITE/IMMULITE 1000 KIT LOT 431, AND FOR IMMULITE 2000 BEGINNING WITH KIT LOT 331. AN URGENT FIELD SAFETY NOTICE (UFSN) 1113 FOR IMMULITE/IMMULITE 1000 AND UFSN 3021 FOR IMMULITE 2000 WAS SENT TO OUS CUSTOMERS IN NOVEMBER OF 2014. THE UFSN RECOMMENDS THAT THE CUSTOMER DISCARD THE AFFECTED KIT LOTS AND TO REVIEW THE LETTER WITH THEIR MEDICAL DIRECTOR.
THE CUSTOMER HAS REQUESTED ADDITIONAL INFORMATION REGARDING METHOD COMPARISON, STUDY ON SAMPLES >5.5 NG/ML, AND THE REFERENCE RANGES FOR THE ANDROSTENEDIONE ASSAY ON THE IMMULITE 2000 INSTRUMENT IN ORDER TO CARRY OUT A METHOD VALIDATION STUDY. IT IS UNKNOWN IF ANY RESULTS ARE BEING REPORTED TO THE PHYSICIAN(S). THERE WERE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 788044 | IMMULITE 2000 ANDROSTENEDIONE | IMMULITE 2000 ANDROSTENEDIONE | CIZ | SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS LIMITED | IMMULITE 1000 ANDROSTENEDIONE | NOT PROVIDED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |