FDA Adverse Event Injury Summary report: N

STRATA® NSC VALVE, REGULAR

MDR report key: 4298882 · Received December 4, 2014

Report

Report Number
2021898-2014-00482
Event Type
Injury
Date Received
December 4, 2014
Date of Event
November 7, 2014
Report Date
November 7, 2014
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K033850
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED VALVE WAS NOT PATENT. PROTEINACEOUS DEBRIS WAS OBSERVED IN THE INTERIOR OF THE VALVE. THE LEAKAGE, REFLUX, PRESSURE-FLOW, AND PREIMPLANTATION TESTS WERE PRECLUDED DUE TO THE VALVE OCCLUSION. THE INSTRUCTIONS FOR USE THAT ACCOMPANIES THE DEVICE STATES THAT THE VALVE MAY BECOME OCCLUDED INTERNALLY DUE TO TISSUE FRAGMENTS, BLOOD CLOTS, TUMOR CELL AGGREGATES, BACTERIAL COLONIZATION OR OTHER DEBRIS. A REVIEW OF THE MANUFACTURING RECORDS SHOWED NO ANOMALIES. (B)(4).

Additional Manufacturer Narrative · 1

THE RETURNED VALVE WAS NOT PATENT. PROTEINACEOUS DEBRIS WAS OBSERVED IN THE INTERIOR OF THE VALVE. THE LEAKAGE, REFLUX, PRESSURE-FLOW, AND PREIMPLANTATION TESTS WERE PRECLUDED DUE TO THE VALVE OCCLUSION. THE INSTRUCTIONS FOR USE THAT ACCOMPANIES THE DEVICE STATES THAT THE VALVE MAY BECOME OCCLUDED INTERNALLY DUE TO TISSUE FRAGMENTS, BLOOD CLOTS, TUMOR CELL AGGREGATES, BACTERIAL COLONIZATION OR OTHER DEBRIS. A REVIEW OF THE MANUFACTURING RECORDS WAS NOT POSSIBLE AS A LOT NUMBER WAS NOT PROVIDED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO MEDTRONIC NEUROSURGERY THAT A STRATA VALVE THAT WAS IMPLANTED SOMETIME IN 2013 WAS EXPLANTED FROM THE PATIENT BECAUSE THE SURGEON SUSPECTED THAT IT WAS OCCLUDED. IT WAS ALSO REPORTED THAT THE PATIENT IS CURRENTLY DOING GREAT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
783033 STRATA® NSC VALVE, REGULAR SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS JXG MEDTRONIC NEUROSURGERY UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 00024 YR Hospitalization| R