ADM RIM IMPACTOR TIP
Report
- Report Number
- 0002249697-2014-04506
- Event Type
- Malfunction
- Date Received
- December 3, 2014
- Date of Event
- November 6, 2014
- Report Date
- November 6, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT REGARDING DAMAGE INVOLVING AN ADM IMPACTOR WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: A MATERIAL ANALYSIS REPORT (MAR) CONCLUDED THAT THE FRACTURE WAS CONSISTENT WITH THE APPLICATION OF OFF-AXIS OVERLOAD STRESSES. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE FRACTURE SURFACES EXAMINED DEVICE HISTORY REVIEW: A DEVICE HISTORY REVIEW INDICATED ALL DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES COMPLAINT HISTORY REVIEW: THERE HAS BEEN NO OTHER COMPLAINTS FOR THE REPORTED LOT. CONCLUSIONS: THIS EVENT WAS PREVIOUSLY INVESTIGATED UNDER INVESTIGATION AND CAPA.
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
IT WAS REPORTED DURING PRIMARY SURGERY, WHEN DR. WAS IMPACTING THE ADM CUP, THE CUP IMPACTOR WOULD NOT HOLD THE CUP TIGHT ENOUGH SO WHEN HE STARTED TO IMPACT THE CUP,THE CUP WOULD MOVE ON THE IMPACTOR AND THEREFORE THROW OFF HIS VERSION AND PLACEMENT OF THE CUP. HE THEN TRIED TO MOVE THE CUP WITH THE SECONDARY RIM IMPACTOR TIP(1235-0-014). WHILE TRYING TO MOVE THE CUP THE RIM IMPACTOR TIP BROKE SO WE HAD TO REMOVE THE ADM CUP AND REPOSITION. ON THE SECOND ATTEMPT TO IMPACT THE CUP THE SAME ISSUE HAPPENED WHERE THE IMPACTOR WOULD NOT HOLD THE CUP TIGHT ENOUGH FOR HIM TO IMPACT THE CUP PROPERLY SO WE BAILED ON THE ADM AND IMPLANTED A TRITANIUM PRIMARY SHELL.
IT WAS REPORTED DURING PRIMARY SURGERY, WHEN DR. WAS IMPACTING THE ADM CUP, THE CUP IMPACTOR WOULD NOT HOLD THE CUP TIGHT ENOUGH SO WHEN HE STARTED TO IMPACT THE CUP,THE CUP WOULD MOVE ON THE IMPACTOR AND THEREFOR THROW OFF HIS VERSION AND PLACEMENT OF THE CUP. HE THEN TRIED TO MOVE THE CUP WITH THE SECONDARY RIM IMPACTOR TIP((B)(4)). WHILE TRYING TO MOVE THE CUP THE RIM IMPACTOR TIP BROKE SO WE HAD TO REMOVE THE ADM CUP AND REPOSITION. ON THE SECOND ATTEMPT TO IMPACT THE CUP THE SAME ISSUE HAPPENED WHERE THE IMPACTOR WOULD NOT HOLD THE CUP TIGHT ENOUGH FOR HIM TO IMPACT THE CUP PROPERLY SO WE BAILED ON THE ADM AND IMPLANTED A TITANIUM PRIMARY SHELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 780710 | ADM RIM IMPACTOR TIP | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS-MAHWAH | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Other |