FDA Adverse Event Injury Summary report: N

MINIBORE PCA EXTENSION SET -67 I

MDR report key: 4284347 · Received November 21, 2014

Report

Report Number
MW5039303
Event Type
Injury
Date Received
November 21, 2014
Date of Event
September 11, 2014
Report Date
November 21, 2014
Manufacturer
HOSPIRA
Product Code
LJS
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

MINIBORE PCA EXTENSION SET 67 INCH. HOSPIRA # (B)(4) WAS PRIMED WITH MORPHINE 1MG/ML FROM PCAIDE ALL THE WAY TO THE OPTION LOCK MALE ADAPTER. THE INSTRUCTIONS ARE CORRECT ON THE PACKAGE TO ONLY PRIME TO THE Y-SITE. THERE IS NO MARKING AT THE Y-SITE TO WARN YOU THAT YOU HAVE REACHED THAT POINT. THUS IT EASY PRIME PAST THIS POINT. ON (B)(6), 2014, 15:30 STOP AT 16:25 MINIBORE PCA EXTENSION SET 67 INCH. HOSPIRA (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
756819 MINIBORE PCA EXTENSION SET -67 I MINIBORE PCA EXTENSION SET -67 I LJS HOSPIRA 14277-28

Patients

Seq Age Sex Outcome Treatment
1 3 MO Hospitalization