FDA Adverse Event
Injury
Summary report: N
MINIBORE PCA EXTENSION SET -67 I
MDR report key: 4284347
·
Received November 21, 2014
Report
- Report Number
- MW5039303
- Event Type
- Injury
- Date Received
- November 21, 2014
- Date of Event
- September 11, 2014
- Report Date
- November 21, 2014
- Manufacturer
- HOSPIRA
- Product Code
- LJS
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
MINIBORE PCA EXTENSION SET 67 INCH. HOSPIRA # (B)(4) WAS PRIMED WITH MORPHINE 1MG/ML FROM PCAIDE ALL THE WAY TO THE OPTION LOCK MALE ADAPTER. THE INSTRUCTIONS ARE CORRECT ON THE PACKAGE TO ONLY PRIME TO THE Y-SITE. THERE IS NO MARKING AT THE Y-SITE TO WARN YOU THAT YOU HAVE REACHED THAT POINT. THUS IT EASY PRIME PAST THIS POINT. ON (B)(6), 2014, 15:30 STOP AT 16:25 MINIBORE PCA EXTENSION SET 67 INCH. HOSPIRA (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 756819 | MINIBORE PCA EXTENSION SET -67 I | MINIBORE PCA EXTENSION SET -67 I | LJS | HOSPIRA | 14277-28 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 MO | Hospitalization |