FDA Adverse Event
Malfunction
Summary report: N
MISTRAL-AIR PLUS (115V) BLOWER
MDR report key: 4284141
·
Received October 1, 2014
Report
- Report Number
- 3003312341-2014-00008
- Event Type
- Malfunction
- Date Received
- October 1, 2014
- Date of Event
- August 21, 2014
- Report Date
- October 1, 2014
- Manufacturer
- THE SURGICAL COMPANY INTERNATIONAL BV DBA THE 37COMPANY
- Product Code
- DWJ
- PMA / PMN Number
- K101705
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL CASE A MISTRAL AIR UNIT STARTED TO PRODUCE SMOKE AND CAUGHT FIRE WHICH WAS IMMEDIATELY EXTINGUISHED. THE HOSE WAS REPORTED TO HAVE BEEN TAPPED TO REPAIR A SPLIT IN THE HEATER HOSE. THE TAPE WAS REPORTED TO HAVE ACTED AS A COMBUSTIBLE MATERIAL WHICH CAUSED THE PLASTIC TO MELT AT THE HOSE AND DEVICE CONNECTION. THE UNIT WAS REMOVED FROM SERVICE IMMEDIATELY AND NO INJURY TO THE PT OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 610203 | MISTRAL-AIR PLUS (115V) BLOWER | THERMAL REGULATING | DWJ | THE SURGICAL COMPANY INTERNATIONAL BV DBA THE 37COMPANY | MA1100-PM |