NON-HEATED ADULT BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00849
- Event Type
- Malfunction
- Date Received
- November 27, 2014
- Date of Event
- October 29, 2014
- Report Date
- October 31, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICES RETURNED: 1 X RT134 LOT 120215 DATE OF MANUFACTURE: 15 FEBRUARY 2012, 1 X RT134 LOT 120306 DATE OF MANUFACTURE: 06 MARCH 2012. METHOD: TWO RT134 NON-HEATED ADULT BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. THEY WERE PRESSURE TESTED AND SUBSEQUENTLY SUBMERGED IN A WATER BATH TO TEST FOR LEAK. RESULTS: THE PRESSURE TEST REVEALED THAT BOTH OF THE RETURNED CIRCUITS EXHIBITED LEAK AND WERE OUT OF SPECIFICATION. THE WATER BATH TEST SHOWED THAT THE LEAK ON BOTH CIRCUITS WAS THROUGH THE CONNECTION BETWEEN THE LID AND BOWL OF THE INSPIRATORY AND EXPIRATORY WATER TRAPS. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBERS 120215 AND 120306. CONCLUSION: THE BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND LID, WHICH CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT ANY LEAK MUST HAVE DEVELOPED AFTER THE BREATHING CIRCUIT WAS RELEASED FOR DISTRIBUTION, DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT134 NON-HEATED ADULT BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." THE HOSPITAL STAFF CORRECTLY CHECKED THE BREATHING CIRCUIT BEFORE PATIENT USE, WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT TWO RT134 ADULT BREATHING CIRCUITS FAILED THE SERVO-S VENTILATOR LEAK TEST. THEY FURTHER STATED THAT THE LEAK OCCURRED FROM THE WATER TRAP.THIS WAS FOUND BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770610 | NON-HEATED ADULT BREATHING CIRCUIT | BZO | BZO | FISHER & PAYKEL HEALTHCARE LTD | RT134 | 120215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SERVO-S VENTILATOR |