GAMBRO WATER PURIFICATION SYSTEM

MDR report key: 4282249 · Received November 26, 2014

Report

Report Number
3019131-2014-00032
Event Type
Malfunction
Date Received
November 26, 2014
Date of Event
October 27, 2014
Report Date
November 26, 2014
Manufacturer
MAR COR PURIFICATION
Product Code
FIP
PMA / PMN Number
K981085
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DUE TO THE WATER SATURATION OF THE ELECTRONICS AND MAIN CWP RO CONTROL PANEL THE CUSTOMER WAS ADVISED TO REPLACE THE ENTIRE CWP RO CONTROL PANEL. PATIENTS WERE NOT ABLE TO DIALYZE UNTIL DEIONIZED WATER TANKS WERE BROUGHT ONLINE, AFTER WHICH ALL PATIENTS WERE ABLE TO BE DIALYZED. THERE WERE NO REPORTS OF PATIENT ILLNESS OR INJURY AS A RESULT OF THIS INCIDENT. MAR COR PURIFICATION WILL CONTINUE TO MONITOR THIS COMPLAINT WITHIN THE COMPLAINT SYSTEM.

Description of Event or Problem · 1

MAR COR PURIFICATION INSTALLED A 10 CUBIC FOOT CARBON TANK IN THE CLINIC'S PRETREATMENT PORTION OF THE REVERSE OSMOSIS WATER SYSTEM IN (B)(6) OF 2013. MCP WAS NOTIFIED ON (B)(6) 2014 THAT THE CARBON TANK FEED SIDE UNION LEAKED AND SPRAYED WATER ON THE CEILING. THE WATER WAS DEFLECTED ONTO THE TOP OF THE CLINICS CWP REVERSE OSMOSIS WATER SYSTEM CAUSING THE ELECTRONICS AND MAIN CONTROL PANEL TO BECOME WET. THIS CAUSED THE CWP TO STOP GENERATING REVERSE OSMOSIS WATER RESULTING IN THE DELAY OF PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
767097 GAMBRO WATER PURIFICATION SYSTEM WATER PURIFICATION SYSTEM FOR HEMODIALYSIS FIP MAR COR PURIFICATION 100474

Patients

Seq Age Sex Outcome Treatment
1