EXETER V40 STEM 44MM NO 2
Report
- Report Number
- 0002249697-2014-04390
- Event Type
- Malfunction
- Date Received
- November 26, 2014
- Date of Event
- October 29, 2014
- Report Date
- October 30, 2014
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- JDI
- PMA / PMN Number
- K011623
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NURSE
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
AN EVENT REGARDING FOREIGN MATTER INVOLVING AN EXERTER STEM WAS REPORTED. THE EVENT WAS CONFIRMED. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: A VISUAL INSPECTION CONFIRMED A HAIR WAS PRESENT FROM THE INSIDE OF THE INNER BLISTER TO THE OUTSIDE OF THE EXTERNAL BLISTER, THROUGH THE SEALING FLANGES. STRYKER (B)(4) PERFORMED AN INTEGRITY TEST (DYE PENETRATION) AND IT WAS VERIFIED THAT THE SEAL INTEGRITY WAS INTACT. DEVICE HISTORY REVIEW: DEVICE HISTORY REVIEW INDICATED ALL DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: A REVIEW OF THE COMPLAINT HISTORY DATABASE SHOWS THAT THERE HAVE BEEN NO SIMILAR REPORTED EVENTS FOR THE SUBJECT LOT CODE. CONCLUSIONS: THE INVESTIGATION CONCLUDED THAT A HAIR WAS PRESENT WITHIN THE SEALS OF THE RETURNED PACKAGING. LMO WAS RAISED BY THE SUPPLIER TO INVESTIGATE FURTHER.
THE SALES REP. HAS REPORTED ON BEHALF OF THE CUSTOMER, THAT DURING SURGERY IT WAS NOTICED THAT A LONG, DARK HAIR WAS SEALED HALF INSIDE THE INNER PACKAGING OF AN EXETER STEM AND HALF OUTSIDE WHICH HAS ALLEGEDLY COMPROMISED THE STERILITY OF THE IMPLANT. THE SALES REP HAS REPORTED THAT ANOTHER STEM WAS IMMEDIATELY AVAILABLE TO COMPLETE THE SURGERY AS THEY KEEP MULTIPLE LEVELS OF STOCK ON THEIR SHELVES.
THE SALES REP. HAS REPORTED ON BEHALF OF THE CUSTOMER, THAT DURING SURGERY IT WAS NOTICED THAT A LONG, DARK HAIR WAS SEALED HALF INSIDE THE INNER PACKAGING OF AN EXETER STEM AND HALF OUTSIDE WHICH HAS ALLEGEDLY COMPROMISED THE STERILITY OF THE IMPLANT. THE SALES REP HAS REPORTED THAT ANOTHER STEM WAS IMMEDIATELY AVAILABLE TO COMPLETE THE SURGERY AS THEY KEEP MULTIPLE LEVELS OF STOCK ON THEIR SHELVES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 768353 | EXETER V40 STEM 44MM NO 2 | IMPLANT | JDI | STRYKER ORTHOPAEDICS-MAHWAH | G4416006 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |