FDA Adverse Event Injury Summary report: N

MAQUET HEATER COOLER UNIT HCU 30

MDR report key: 4280275 · Received November 26, 2014

Report

Report Number
8010762-2014-01217
Event Type
Injury
Date Received
November 26, 2014
Date of Event
June 21, 2012
Report Date
June 23, 2012
Manufacturer
MAQUET CARDIOPULMOARY AG
Product Code
DWC
PMA / PMN Number
K031544
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PER SERVICE REPORT (B)(4), THE REPORTED EVENT WAS NOT REPRODUCED, HOWEVER, THE 3 WAY VALVE AND VALVE ACTUATOR WERE REPLACED. THE UNIT WAS TESTED TO FACTORY SPECIFICATIONS AND PASSED ALL TESTS. THE REPLACED PARTS WERE SENT TO MCP (B)(4) FOR EVALUATION AND PER ANALYSIS REPORT (B)(4) THE ACTUATOR AND VALVE WERE FUNCTIONING CORRECTLY, HOWEVER, DURING THE ASSEMBLY OF THE HCU 30 3 DISTANCE WASHERS WERE NEEDED IN THE ADAPTOR HOUSING. THIS IS REQUIRED TO LOOSEN THE CONTACT BETWEEN THE BUNG OF THE EMO 1 AND THE BUNG OF THE 3 WAY VALVE. (B)(4).

Description of Event or Problem · 1

ON (B)(6) 2012 AT (B)(6), IT WAS REPORTED THE HEATER COOLER UNIT 30 (HCU 30) SERIAL NUMBER (B)(4) WOULD NOT REWARM THE PATIENT. THE PERFUSIONIST HAD TO KEEP THE PATIENT ON BYPASS FOR AN ADDITIONAL 20 MINUTES WHILE THE DEVICE WAS REPLACED WITH ANOTHER. THE PATIENT WAS STABLE POST COMPLETION OF THE PROCEDURE. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
768109 MAQUET HEATER COOLER UNIT HCU 30 CONTROLLER, TEMPERATURE, CPB DWC MAQUET CARDIOPULMOARY AG 70103.4371

Patients

Seq Age Sex Outcome Treatment
1 Congenital Anomaly