FDA Adverse Event
Malfunction
Summary report: N
G.E. ADVANTX
MDR report key: 42791
·
Received October 10, 1996
Report
- Report Number
- MW1010045
- Event Type
- Malfunction
- Date Received
- October 10, 1996
- Date of Event
- September 30, 1996
- Report Date
- September 30, 1996
- Manufacturer
- GENERAL ELECTRIC
- Product Code
- KXJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MN, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
WHILE RAISING THE X-RAY TABLE FROM THE HORIZONTAL TO VERTICAL POSITION, THE TABLE SLIPPED OFF ITS PIVOT JOINT AND DROPPED 4-6 INCHES TO THE FLOOR. IT WAS NOTED THAT A BLANKET WHICH HAD BEEN PLACED UNDER THE PT WAS CAUGHT IN THE TILT MECHANISM. ACCORDING TO THE CO SERVICEMEN, THE BLANKET HAD WRAPPED AROUND THE SPROCKET CAUSING THE CHAIN TO COME OFF THE SPROCKET AND WHEN THAT HAPPENED, THE TABLE SLIPPED DOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | G.E. ADVANTX | X-RAY TABLE | KXJ | GENERAL ELECTRIC | 46-276420G1 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Other |