FDA Adverse Event Malfunction Summary report: N

G.E. ADVANTX

MDR report key: 42791 · Received October 10, 1996

Report

Report Number
MW1010045
Event Type
Malfunction
Date Received
October 10, 1996
Date of Event
September 30, 1996
Report Date
September 30, 1996
Manufacturer
GENERAL ELECTRIC
Product Code
KXJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MN, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

WHILE RAISING THE X-RAY TABLE FROM THE HORIZONTAL TO VERTICAL POSITION, THE TABLE SLIPPED OFF ITS PIVOT JOINT AND DROPPED 4-6 INCHES TO THE FLOOR. IT WAS NOTED THAT A BLANKET WHICH HAD BEEN PLACED UNDER THE PT WAS CAUGHT IN THE TILT MECHANISM. ACCORDING TO THE CO SERVICEMEN, THE BLANKET HAD WRAPPED AROUND THE SPROCKET CAUSING THE CHAIN TO COME OFF THE SPROCKET AND WHEN THAT HAPPENED, THE TABLE SLIPPED DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 G.E. ADVANTX X-RAY TABLE KXJ GENERAL ELECTRIC 46-276420G1 NA

Patients

Seq Age Sex Outcome Treatment
1 85 YR Other