SHOECOVER DURA FIT XL 200/PR
Report
- Report Number
- 1423537-2014-00031
- Event Type
- Injury
- Date Received
- November 25, 2014
- Date of Event
- October 30, 2014
- Report Date
- November 25, 2014
- Manufacturer
- CARDINAL HEALTH SINGAPORE 225 PTE LTD
- Product Code
- FXP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO EXCEPTION WAS RECORDED DURING PRODUCTION OF THIS LOT. THE DHR INDICATED THAT THE COEFFICIENT OF FRICTION WAS MEETING THE SPECIFICATION WHEN IT WAS FIRST PRODUCED, IT WAS NOTED THAT THE COF WAS FOUND TO DETERIORATE THROUGH TIME USUALLY IN STORAGE. THIS APPEARS TO BE AN INHERENT CHARACTERISTIC OF THE SHOE COVER. THE ROOT CAUSE FOR THE DETERIORATION CANNOT BE IDENTIFIED AT THIS TIME. CURRENTLY, THERE ARE NO DOCUMENTED CRITERIA FOR COEFFICIENT OF FRICTION. AS A RESULT, CARDINAL HEALTH HAS UTILIZED THE OSHA GUIDELINES FOR FLOOR CONDITIONS TO BE SLIP-RESISTANT AND HAVE A STATIC COEFFICIENT OF FRICTION (SCOF) VALUE OF 0.5 OR GREATER. CARDINAL HEALTH HAS AN INTERNAL COEFFICIENT FRICTION OF 0.68. HISTORICAL TRENDING WAS DONE AND THIS IS THE SECOND REPORT OF SLIPPERINESS IN THE PAST TWELVE MONTHS FOR THIS CATALOG NUMBER. THE COMPLAINT WAS FED BACK TO THE RELEVANT SECTION OF PRODUCTION FOR AWARENESS. A REFRESHER TRAINING SESSION WAS CONDUCTED FOR THE OPERATORS TO ENSURE THEY STRICTLY COMPLY WITH THE SPECIFICATION AND PRODUCTION METHODS. A CAPA PROJECT WAS RAISED TO ANALYZE THE CAUSE OF THE DETERIORATION AND THE ACTION NEEDED TO CONTAIN AND CORRECT THE ISSUE. AT THIS TIME, NO ADDITIONAL CORRECTIVE ACTIONS WILL BE TAKEN, BUT WE WILL CONTINUE TO MONITOR COMPLAINTS FOR TRENDS OF THIS ISSUE.
WHILE WEARING THE SHOE COVERS, A NURSE SLIPPED AND CAUGHT HERSELF, BUT IN THE PROCESS WRENCHED/STRAINED HER BACK AND HAD PAIN FOR SEVERAL HOURS AFTERWARDS. SHE TOOK ADVIL AND APPLIED HEAT. SHE IS FINE NOW AND HAS NO COMPLAINTS AT THIS TIME. THE FLOOR HAD NOT BEEN RECENTLY WASHED/WAXED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 764592 | SHOECOVER DURA FIT XL 200/PR | COVER, SHOE, OPERATING-ROOM | FXP | CARDINAL HEALTH SINGAPORE 225 PTE LTD | 14GEF182 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Other |