FDA Adverse Event Injury Summary report: N

SHOECOVER DURA FIT XL 200/PR

MDR report key: 4276942 · Received November 25, 2014

Report

Report Number
1423537-2014-00030
Event Type
Injury
Date Received
November 25, 2014
Date of Event
October 29, 2014
Report Date
November 25, 2014
Manufacturer
CARDINAL HEALTH SINGAPORE 225 PTE LTD
Product Code
FXP
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO EXCEPTION WAS RECORDED DURING PRODUCTION OF THIS LOT. THE DHR INDICATED THAT THE COEFFICIENT OF FRICTION WAS MEETING THE SPECIFICATION WHEN IT WAS FIRST PRODUCED, IT WAS NOTED THAT THE COF WAS FOUND TO DETERIORATE THROUGH TIME USUALLY IN STORAGE. THIS APPEARS TO BE AN INHERENT CHARACTERISTIC OF THE SHOE COVER. THE ROOT CAUSE FOR THE DETERIORATION CANNOT BE IDENTIFIED AT THIS TIME. CURRENTLY, THERE ARE NO DOCUMENTED CRITERIA FOR COEFFICIENT OF FRICTION. AS A RESULT, CARDINAL HEALTH HAS UTILIZED THE OSHA GUIDELINES FOR FLOOR CONDITIONS TO BE SLIP-RESISTANT AND HAVE A STATIC COEFFICIENT OF FRICTION (SCOF) VALUE OF 0.5 OR GREATER. CARDINAL HEALTH HAS AN INTERNAL COEFFICIENT FRICTION OF 0.68. HISTORICAL TRENDING WAS DONE AND THIS IS THE FIRST REPORT OF SLIPPERINESS IN THE PAST TWELVE MONTHS FOR THIS CATALOG NUMBER. THE COMPLAINT WAS FED BACK TO THE RELEVANT SECTION OF PRODUCTION FOR AWARENESS. A REFRESHER TRAINING SESSION WAS CONDUCTED FOR THE OPERATORS TO ENSURE THEY STRICTLY COMPLY WITH THE SPECIFICATION AND PRODUCTION METHODS. A CAPA PROJECT WAS RAISED TO ANALYZE THE CAUSE OF THE DETERIORATION AND THE ACTION NEEDED TO CONTAIN AND CORRECT THE ISSUE. AT THIS TIME, NO ADDITIONAL CORRECTIVE ACTIONS WILL BE TAKEN, BUT WE WILL CONTINUE TO MONITOR COMPLAINTS FOR TRENDS OF THIS ISSUE.

Description of Event or Problem · 1

WHILE WEARING THE SHOE COVERS, A NURSE FELL AND CRUSHED HER ELBOW. SHE TURNED TO THE RIGHT AND HER FEET FLEW OUT FROM UNDER HER TO THE LEFT. THE FLOOR WAS DRY. SHE REQUIRED SURGERY AND IS EXPECTED TO BE OFF WORK FOR SEVERAL WEEKS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764207 SHOECOVER DURA FIT XL 200/PR COVER, SHOE, OPERATING-ROOM FXP CARDINAL HEALTH SINGAPORE 225 PTE LTD 14GEF182

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R