PRECISION®
Report
- Report Number
- 3006630150-2014-02739
- Event Type
- Malfunction
- Date Received
- November 24, 2014
- Date of Event
- September 28, 2014
- Report Date
- October 28, 2014
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL#: SC-8120-70, SERIAL #: (B)(4), DESCRIPTION: ARTISAN 2X8 SURGICAL LEAD - 70 CM.
SC-1110 (SN (B)(4)) DEVICE EVALUATION INDICATED THAT THE IPG PASSED VISUAL, ELECTRICAL, PERFORMANCE AND PHOTOGRAPHIC IMAGING TESTS PERFORMED. THE COMPLAINT WAS NOT CONFIRMED. DEVICE EXHIBITED NORMAL CHARGING AND DISCHARGING CHARACTERISTICS WHEN CHARGED WITH RECOMMENDED OPTIMAL ALIGNMENT. WHEN ALIGNMENT OPTIMIZATION IS REDUCED, CHARGING COULD LENGTHEN CONSIDERABLY. THE BATTERY WAS DEPLETING ITS CHARGE AT A RATE OF 12 MVDC PER DAY WITH THE STIMULATION TURNED OFF, WHICH WAS WITHIN THE TYPICAL IPG BATTERY DEPLETION RATE. THE COMPLAINT OF HIGH IMPEDANCE WAS NOT CONFIRMED AS VARIOUS TEST LEADS WERE INSERTED IN BOTH PORTS. ALL IMPEDANCE READINGS WERE WITHIN NORMAL RANGE. HIGH IMPEDANCE READINGS COULD NOT BE DUPLICATED IN THE LAB. THE DEVICE EXHIBITED NORMAL DEVICE CHARACTERISTICS. SC-8120-70 (SN (B)(4)) THE COMPLAINT OF HIGH IMPEDANCE ON CONTACT # 6 WAS CONFIRMED. VISUAL (MICROSCOPE) AND X-RAY INSPECTION FOUND FRACTURED CABLES AT THE BENT/KINKED LOCATION OF THE LEAD. IT APPEARED TO BE WHERE THE LEAD BODIES WERE SUTURED. HIGH IMPEDANCE READINGS WERE REGISTERED ON CONTACTS # 1, 6, 7, 8, AND 14. THE LEAD BODY OF RIGHT PIGTAIL SHOWED EXPOSED CABLES AND MINOR CUTS. IN ADDITION, THE PADDLE WAS PARTIALLY TORN RIGHT AT THE JUNCTION OF THE TWO LEAD BODIES. THIS PADDLE DAMAGE WAS A RESULT OF A TYPICAL EXPLANT PROCEDURE AND NOT CONSIDERED A FAILURE. THE HIGH IMPEDANCE COMPLAINT ON CONTACTS # 9 AND 15 WERE NOT CONFIRMED.
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT UNDERWENT AN IPG AND LEAD REPLACEMENT. THE PHYSICIAN BELIEVED THERE WAS NO MALFUNCTION ON THE IPG AND STATED THAT LEAD TAIL MAY HAVE BEEN DAMAGED WHERE THE SUTURE SLEEVE WAS.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING FREQUENT IPG CHARGING. DATABASE ANALYSIS REVEALED HIGH IMPEDANCES ON FEW CONTACTS OF THE LEAD AND THE BATTERY DISCHARGE WAS OBSERVED TO BE WITHIN AN EXPECTED RANGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE AND POSSIBLE REPLACEMENT OF THE LEAD.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING FREQUENT IPG CHARGING. DATABASE ANALYSIS REVEALED HIGH IMPEDANCES ON FEW CONTACTS OF THE LEAD AND THE BATTERY DISCHARGE WAS OBSERVED TO BE WITHIN AN EXPECTED RANGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE AND POSSIBLE REPLACEMENT OF THE LEAD.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING FREQUENT IPG CHARGING. DATABASE ANALYSIS REVEALED HIGH IMPEDANCES ON FEW CONTACTS OF THE LEAD AND THE BATTERY DISCHARGE WAS OBSERVED TO BE WITHIN AN EXPECTED RANGE. THE PATIENT WILL UNDERGO AN IPG REPLACEMENT PROCEDURE AND POSSIBLE REPLACEMENT OF THE LEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760172 | PRECISION® | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-1110 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |