FDA Adverse Event
Malfunction
Summary report: N
21G HOFFMAN/AHMED HORIZONTAL SCISSOR
MDR report key: 4268543
·
Received November 4, 2014
Report
- Report Number
- 3019924-2014-00050
- Event Type
- Malfunction
- Date Received
- November 4, 2014
- Report Date
- October 30, 2014
- Manufacturer
- MICROSURGICAL TECHNOLOGY, INC.
- Product Code
- HNF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VE
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED WITH CORROSION AND PITTING AT THE SITE OF THE FRACTURE WHICH INDICATES THE DEVICE WAS NOT CLEANED AND MAINTAINED ADEQUATELY.
Description of Event or Problem · 1
THE SURGEON REPORTED THAT WHILE USING A HOFFMAN/AHMED HORIZONTAL SCISSORS TO CUT IRIS TISSUE THE SCISSOR BLADE BROKE. THE BROKEN PIECE WAS REMOVED FROM THE EYE AND THERE WAS NO IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 706672 | 21G HOFFMAN/AHMED HORIZONTAL SCISSOR | OPTHALMIC SCISSOR | HNF | MICROSURGICAL TECHNOLOGY, INC. | DFH-0005 | 049444 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |