FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4261404 · Received November 19, 2014

Report

Report Number
2032227-2014-53906
Event Type
Malfunction
Date Received
November 19, 2014
Date of Event
October 21, 2014
Report Date
October 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE, DUE TO FLATTENED ESC AND ACT BUTTON DOME SWITCHES. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE ON DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAS AN UNRESPONSIVE KEYPAD. CUSTOMER'S BLOOD GLUCOSE READING WAS 180 MG/DL. NO SIGNIFICANT EVENTS LEADING TO THE ALARM WERE OBSERVED. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
751037 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 66 YR