FDA Adverse Event
Malfunction
Summary report: N
GEMSTAR 7 THPY ALT ST
MDR report key: 4260747
·
Received September 4, 2014
Report
- Report Number
- 9615050-2014-05120
- Event Type
- Malfunction
- Date Received
- September 4, 2014
- Date of Event
- August 16, 2014
- Report Date
- August 19, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1169-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND LEAKAGE FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT. AS INDICATED, THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A CUSTOMER CONTACT REPORT THAT DURING POWER UP THE PUMP ALARMED FOR ERROR CODE 11/004. THIS DOES NOT INDICATE A REPORTABLE EVENT. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 540899 | GEMSTAR 7 THPY ALT ST | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |