FDA Adverse Event Malfunction Summary report: N

GEMSTAR 7 THPY ALT ST

MDR report key: 4260747 · Received September 4, 2014

Report

Report Number
9615050-2014-05120
Event Type
Malfunction
Date Received
September 4, 2014
Date of Event
August 16, 2014
Report Date
August 19, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1169-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND LEAKAGE FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT. AS INDICATED, THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A CUSTOMER CONTACT REPORT THAT DURING POWER UP THE PUMP ALARMED FOR ERROR CODE 11/004. THIS DOES NOT INDICATE A REPORTABLE EVENT. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
540899 GEMSTAR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA