FDA Adverse Event Malfunction Summary report: N

ON-X CONFORM-X AORTIC PROSTHETIC HEART VALVE

MDR report key: 4260745 · Received November 14, 2014

Report

Report Number
1649833-2014-00014
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
September 26, 2014
Report Date
November 14, 2014
Manufacturer
ON-X LIFE TECHNOLOGIES, INC.
Product Code
LWQ
PMA / PMN Number
P000037
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VALVE HAS BEEN RETURNED, INVESTIGATION NOT YET COMPLETED. IN ALL PAST CASES OF CARBON PART BREAKAGE, THE INVESTIGATION CONCLUSION WAS "IATROGENIC", IE, SOME SORT OF IMPROPER, ROUGH TREATMENT OF THE VALVE. THERE HAS BEEN NO STRUCTURAL DYSFUNCTION OF FAILURE OF THE ON-X PROSTHETIC OF THE INVESTIGATION IS EXPECTED TO BE SIMILAR TO THAT IN THE PAST. WHEN RELEVANT FURTHER INFORMATION BECOMES AVAILABLE, A FOLLOW-UP MDR WILL BE FILED.

Description of Event or Problem · 1

DURING IMPLANTATION, A 2 MM PIECE OF THE CARBON HOUSING WAS BROKEN. THE 2 MM PIECE WAS REMOVED. THE VALVE WAS EXPLANTED AND A NEW/DIFFERENT VALVE WAS IMPLANTED IN ITS PLACE. THE 2 MM PIECE WAS DISCARDED AT THE HOSPITAL. THE VALVE HOUSING IS BEING INVESTIGATED FOR FRACTURE ANALYSIS AT AN INDEPENDENT LAB TO DETERMINE THE ORIGIN AND POTENTIAL CAUSE OF THE CRACK. THIS IS AN INITIAL REPORT. CARBON PART BREAKAGE IS SEPARATELY TRACKED TO ENSURE NO TRENDS OR SYSTEMIC ASSIGNABLE CAUSE. CURRENTLY, ALL HAVE OCCURRED INTRA-OPERATIVELY AND ROOT CAUSE FOR ALL HAVE BEEN KNOWN OR SHOWN TO BE IATROGENIC IN NATURE, AND NOT DUE TO PRODUCT DEFECT OR MALFUNCTION. SUCH BREAKAGE HAS OCCURRED AT A RECURRENCE RATE OF ABOUT 65 PPM. THERE WILL BE A FOLLOW UP AFTER MANUFACTURER'S INVESTIGATION. PER DISTRIBUTOR-PROVIDED INFORMATION, THE PATIENT IS IN GOOD CONDITION AFTER THE SURGERY, WITHOUT ANY COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
738826 ON-X CONFORM-X AORTIC PROSTHETIC HEART VALVE MECHANICAL HEART VALVE LWQ ON-X LIFE TECHNOLOGIES, INC. ONXACE

Patients

Seq Age Sex Outcome Treatment
1 UNK Other