PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-54023
- Event Type
- Malfunction
- Date Received
- November 18, 2014
- Date of Event
- March 4, 2014
- Report Date
- October 22, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP AND A MISSING END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
IT WAS REPORTED THAT MOTOR ERROR AND PRIME REWIND ALARMS WERE RECEIVED ON THE CUSTOMER'S INSULIN PUMP. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARED NORMAL. THE INSULIN PUMP HAD NOT RECENTLY BEEN BUMPED OR DROPPED PRIOR TO THE ALARM OCCURRENCE, BUT HAD BEEN DROPPED IN THE PAST. THE CUSTOMER DECLINED TO TROUBLESHOOT FOR THE PRIME REWIND ALARM. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 744270 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |