FDA Adverse Event
Injury
Summary report: N
TALENT AAA STENT GRAFT
MDR report key: 4260222
·
Received November 18, 2014
Report
- Report Number
- 2953200-2014-02425
- Event Type
- Injury
- Date Received
- November 18, 2014
- Date of Event
- November 10, 2014
- Report Date
- November 10, 2014
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P070027
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
A TALENT AAA STENT GRAFT SYSTEM WAS IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF A 70 CM DIAMETER ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT CAME IN FOR ROUTINE CT FOR SURVEILLANCE POST ENDOVASCULAR ANEURYSM REPAIR (EVAR) AND IT WAS DISCOVERED THAT A PREVIOUSLY IMPLANTED STENT GRAFT HAD MIGRATED AND HAD A TYPE IA ENDOLEAK INTO AN ENLARGED AORTIC ABDOMINAL ANEURYSM. A CUFF WAS SUCCESSFULLY PLACED TO SEAL OFF THE TYPE IA ENDOLEAK AND THE ISSUE RESOLVED. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE ROOT CAUSE OF THE MIGRATION AND TYPE IA ENDOLEAK WAS DISEASE PROGRESSION. NO CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 746821 | TALENT AAA STENT GRAFT | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00079 YR | Required Intervention |