PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-53884
- Event Type
- Malfunction
- Date Received
- November 18, 2014
- Date of Event
- October 21, 2014
- Report Date
- October 21, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED CASE ON THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED BELT CLIP SLOT.
IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP HAD NONRESPONSIVE "ACT" AND "UP" BUTTONS. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 145 MG/DL. STANDARD TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE PRODUCT HAS BEEN RETURNED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 746818 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |