FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4260212 · Received November 18, 2014

Report

Report Number
2032227-2014-53881
Event Type
Malfunction
Date Received
November 18, 2014
Date of Event
October 21, 2014
Report Date
October 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED A33 DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME/A33 TEST DUE TO PROTRUDED LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A COMPROMISED FORCE SENSOR SYSTEM ERROR FROM HIS INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 243 MG/DL. THE CUSTOMER STATED THE ALARM OCCURRED DURING PRIME AND THAT HE ATTEMPTED TO PRIME THREE TIMES WITHOUT SUCCESS. STANDARD TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. THE PRODUCT HAS BEEN RETURNED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
746239 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 31 YR