ENDURANT II
Report
- Report Number
- 2953200-2014-02424
- Event Type
- Malfunction
- Date Received
- November 18, 2014
- Date of Event
- November 12, 2014
- Report Date
- November 12, 2014
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- MIH
- PMA / PMN Number
- P100021
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). EVALUATION, CONCLUSION: OFF-LABEL (NECK ANGULATION GREATER THAN 60 DEGREE, NECK LENGTH LESS THAN 10MM).
AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF A 6 CM DIAMETER ABDOMINAL AORTIC ANEURYSM. THE PROXIMAL NECK WAS 23 MM IN DIAMETER AND 7MM IN LENGTH. THE LEFT COMMON ILIAC ARTERY WAS 9 MM IN DIAMETER AND THE RIGHT COMMON ILIAC ARTERY WAS 20 MM IN DIAMETER. THE RIGHT AND LEFT EXTERNAL ILIAC ARTERIES MEASURED 7.6 MM IN DIAMETER. THE AORTIC NECK ANGULATION WAS 70 DEGREES. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE THE BIFURCATE WAS INSERTED FROM THE RIGHT SIDE HOWEVER, THE PHYSICIAN HAD DIFFICULTY IN LANDING TO THE GREATER CURVATURE OF PROXIMAL NECK. SO, THE MAIN BODY WAS INSERTED FROM LEFT SIDE. THERE WAS NO SUFFICIENT SEALING SPACE, BUT THE STENT GRAFT WAS SUCCESSFULLY DEPLOYED WITH PUSH-UP. INITIAL TOUCH-UP WAS NOT CARRIED OUT AS THE PROXIMAL AORTIC NECK DIDN¿T HAVE ENOUGH TOUCH-UP SPACE. ON THE FINAL ANGIOGRAM, A PROXIMAL TYPE I ENDOLEAK WAS OBSERVED SO THE PHYSICIAN BALLOONED HOWEVER, THE ENDOLEAK REMAINED. THE PHYSICIAN COMMENTED THAT ¿IT WAS QUITE A DIFFICULT CASE. IN CASE OF NORMAL CASE, OPEN SURGERY WOULD BE PERFORMED. BUT THE PATIENT HAD OPEN SURGERY IN THE PAST SO THAT ANOTHER OPEN SURGERY SHOULD BE AVOIDED.¿ NO ADDITIONAL TREATMENT IS PLANNED. NO CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS BEING MONITORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 746182 | ENDURANT II | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | MEDTRONIC IRELAND | V05904373 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00077 YR |