PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-54503
- Event Type
- Injury
- Date Received
- November 18, 2014
- Date of Event
- August 9, 2014
- Report Date
- October 21, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. REFERENCE MANUFACTURER REPORT NUMBER: 2032227-2014-54504.
IT WAS REPORTED THE CUSTOMER WAS RECEIVING NO DELIVERY ALARMS. CUSTOMER ALSO REPORTED THEIR BLOOD GLUCOSE REMINDER WAS ALSO NOT FUNCTIONING. THE CUSTOMER ALSO REPORTED HAVING A LOW BLOOD GLUCOSE EVENT WHERE THEIR BLOOD GLUCOSE DECLINED TO 24 MG/DL. THE CUSTOMER'S BLOOD GLUCOSE WAS 175 MG/DL AT THE TIME OF THE CALL. THE CUSTOMER ALSO REPORTED A MOTOR ERROR ALARM THAT OCCURRED DURING A BOLUS DELIVERY. CUSTOMER ALSO STATED THEY WERE ABLE TO COMPLETE THE REWIND SEQUENCE ON THEIR INSULIN PUMP. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 746450 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |