CAPTURE-R READY INDICATOR RED CELLS
Report
- Report Number
- 1034569-2014-00207
- Event Type
- Malfunction
- Date Received
- November 18, 2014
- Date of Event
- October 20, 2014
- Report Date
- November 17, 2014
- Manufacturer
- IMMUCOR, INC.
- Product Code
- KSZ
- PMA / PMN Number
- BK020053
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
IMMUCOR TECHNICAL SUPPORT USED A REMOTE ELECTRONIC CONNECTION METHOD TO ASSESS THE INSTRUMENT TEST WELL IMAGES. BATCH 27350 TESTED ON (B)(6) 2014 USING LOTS R474 IC 221242; NEGATIVE REACTIONS FOR CELLS 1,2,3 VISUALLY NEGATIVE. CAPTURE-R READY INDICATOR RED CELL (CRRIRC), LOT 221242, EXPIRED PRIOR TO COMPLAINT BEING REFERRED TO PI. PRODUCT WAS NOT USED ON THE ECHO WHILE IN DATE. DHR REVIEW: -- SCREEN ASSAY WAS PERFORMED ON (B)(6) 2014 USING CRRS3 LOT R456 ON ECHO INSTRUMENT; -- CONTROLS PERFORMED AS EXPECTED. WBCORQC LEVEL 2 CONTAINS ANTI-D. RESULTS WERE 4+ WITH CELL 1, 4+ WITH CELL 2, AND NEGATIVE WITH CELL 3. CELLS 1 AND 2 ARE D ANTIGEN POSITIVE. CELL 3 IS D ANTIGEN NEGATIVE. -- 4 REPLICATES OF CAPTURE-R NEGATIVE CONTROL, LOT 245140L, RESULTED NEGATIVE AS EXPECTED. -- 4 REPLICATES OF CAPTURE-R POSITIVE CONTROL (WEAK), LOT 244290, RESULTED POSITIVE AS EXPECTED. -- 4 REPLICATES OF KNOWN ANTIBODY NEGATIVE (INERT) IN-HOUSE DONORS RESULTED NEGATIVE AS EXPECTED. CRRIRC LOT 221242 PERFORMED AS EXPECTED. DHR QUALITY REVIEW DEEMED ALL SPECIFICATIONS AS BEING MET AND RELEASED PRODUCT TO MARKET ON 12SEP2014.
ON (B)(6) 2014, A CUSTOMER REPORTED AN UNEXPECTED NEGATIVE ANTIBODY SCREEN ASSAY WHEN TESTING A PATIENT SAMPLE USING CAPTURE-R READY INDICATOR RED CELLS (CRRIRC) ON A GALILEO ECHO INSTRUMENT. THE PATIENT IS O NEGATIVE AND HAS A KNOWN ANTI D.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 744711 | CAPTURE-R READY INDICATOR RED CELLS | ANTIGLOBULIN COATED RED CELLS | KSZ | IMMUCOR, INC. | 221242 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |