FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 4256907 · Received November 18, 2014

Report

Report Number
2939301-2014-30996
Event Type
Malfunction
Date Received
November 18, 2014
Report Date
November 4, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOWUP#1 11/18/2014 - CORRECTION:THE FOLLOWING STATEMENT WAS NOT SUMBITTED WITH THE INITAIL REPORT."LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT."

Description of Event or Problem · 1

ON (B)(6) 2014 THE REPORTER CONTACTED LIFESCAN USA ALLEGING INACCURATE LOW CONTROL SOLUTION RESULTS. THE USER IN REPORTED THAT THE QUALITY CONTROL TESTS FAILED SPECIFICATION BECAUSE THE RESULTS FELL BELOW THE SPECIFIED RANGE ON THE TEST STRIP VIAL. THE REPORTER CLAIMED OBTAINING CONTROL SOLUTION READING(S) OF ¿114 MG/DL¿ WITH THE CONTROL SOLUTION FOR THE SUBJECT LOT BEING 122-162 MG/DL. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S ACCURACY CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
745485 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3631984

Patients

Seq Age Sex Outcome Treatment
1