FDA Adverse Event Other Summary report: N

HEWLETT PACKARD PAGE WRITER XLI

MDR report key: 425484 · Received October 29, 2002

Report

Report Number
MW1026578
Event Type
Other
Date Received
October 29, 2002
Date of Event
May 9, 2002
Report Date
October 29, 2002
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
DPS
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
MT, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PT COMPLAINING OF CHEST PAIN. STAT EKG ORDERED. THE EKG SHOWED A WIDE QRS ON THE 12 LEAD EKG. CARDIOLOGY CONSULTED. HEART CATH PERFORMED WITH NO INTERVENTIONS NEEDED. UPON RE-REVIEW OF THE EKG POST-PRODEDURE IT WAS NOTED THAT THE EKG WAS DONE AT 50 MM/SEC INSTEAD OF 25 MM/SEC. THE BUTTON ON THE LEFT SIDE OF THE PANEL THAT CONTROLS THE SPEED WAS INADVERTENTLY BUMPED DURING PAPER CHANGE. THIS CAUSED THE SPEED TO INCREASE FROM 25 MM/SEC TO 50 MM/SEC. THESE CONTROL BUTTONS DO NOT HAVE A PROTECTIVE COVER. ADDITIONALLY THE PROCESS TO CHANGE THE SPEED IS A ONE-STEP PROCESS. DEVICE MAY BE EVALUATED ON SITE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEWLETT PACKARD PAGE WRITER XLI EKG MACHINE DPS PHILIPS MEDICAL SYSTEMS M1700A *

Patients

Seq Age Sex Outcome Treatment
1 55 YR Other