HEWLETT PACKARD PAGE WRITER XLI
MDR report key: 425484
·
Received October 29, 2002
Report
- Report Number
- MW1026578
- Event Type
- Other
- Date Received
- October 29, 2002
- Date of Event
- May 9, 2002
- Report Date
- October 29, 2002
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- DPS
- Adverse Event
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MT, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PT COMPLAINING OF CHEST PAIN. STAT EKG ORDERED. THE EKG SHOWED A WIDE QRS ON THE 12 LEAD EKG. CARDIOLOGY CONSULTED. HEART CATH PERFORMED WITH NO INTERVENTIONS NEEDED. UPON RE-REVIEW OF THE EKG POST-PRODEDURE IT WAS NOTED THAT THE EKG WAS DONE AT 50 MM/SEC INSTEAD OF 25 MM/SEC. THE BUTTON ON THE LEFT SIDE OF THE PANEL THAT CONTROLS THE SPEED WAS INADVERTENTLY BUMPED DURING PAPER CHANGE. THIS CAUSED THE SPEED TO INCREASE FROM 25 MM/SEC TO 50 MM/SEC. THESE CONTROL BUTTONS DO NOT HAVE A PROTECTIVE COVER. ADDITIONALLY THE PROCESS TO CHANGE THE SPEED IS A ONE-STEP PROCESS. DEVICE MAY BE EVALUATED ON SITE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEWLETT PACKARD PAGE WRITER XLI | EKG MACHINE | DPS | PHILIPS MEDICAL SYSTEMS | M1700A | * |