FDA Adverse Event Malfunction Summary report: N

SIG FEM ADPT TORQUE WRENCH

MDR report key: 4253908 · Received November 17, 2014

Report

Report Number
1818910-2014-31752
Event Type
Malfunction
Date Received
November 17, 2014
Date of Event
November 6, 2014
Report Date
November 6, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMED THE BLACK PLASTIC PROTECTOR COMPONENT HAS CRACKED AND BECOME DISASSEMBLED FROM THE WRENCH. CAPA (B)(4) WAS INITIATED TO FURTHER INVESTIGATE AND DETERMINE ROOT CAUSE AND CORRECTIVE ACTIONS. CO 103025887 HAS BEEN INITIATED TO CHANGE THE DESIGN OF PRODUCT CODE 961673 AS PART OF CAPA (B)(4). THE CURRENT COMPLAINT SAMPLE WAS MANUFACTURED PRIOR TO THE IMPLEMENTATION OF CO 103025887. NO FURTHER CORRECTIVE ACTION REQUIRED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

THE TORQUE WRENCH CONNECTION IS WORN.