FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE PRO
MDR report key: 4253530
·
Received November 15, 2014
Report
- Report Number
- 1416980-2014-40841
- Event Type
- Malfunction
- Date Received
- November 15, 2014
- Date of Event
- August 23, 2014
- Report Date
- October 21, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE HIGH DRAIN ALARM WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. A VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED AND NO ISSUES WERE OBSERVED. THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 104 (NIGHT DRAIN 4) ALARM WAS IDENTIFIED IN THE LOG. THIS INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE. THE ALARM OCCURRED ON (B)(6) 2014 AT 07:54:50. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 740720 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |