FDA Adverse Event Malfunction Summary report: N

SYMBIQ DUAL CHANNEL

MDR report key: 4253261 · Received November 12, 2014

Report

Report Number
9615050-2014-06023
Event Type
Malfunction
Date Received
November 12, 2014
Date of Event
January 1, 2014
Report Date
November 6, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, THE DEVICE DID NOT ALARM WITH A S233 (OVERTEMPERATURE-PSC) MALFUNCTION CODE; HOWEVER, IT WAS NOTED IN THE DEVICE HISTORY MULTIPLE TIMES AND THE COOLING FAN DID NOT FUNCTION. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM CODE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT WITH AN UNSIGNED NOTE THAT INDICATED THAT THE DEVICE ALARMS MALFUNCTION S233. NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, MULTIPLE S233 MALFUNCTION ALARM CODES WERE NOTED AND THE COOLING FAN DID NOT FUNCTION. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
729468 SYMBIQ DUAL CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK