HEARTMATE II SYSTEM CONTROLLER
Report
- Report Number
- 2916596-2014-02080
- Event Type
- Malfunction
- Date Received
- November 13, 2014
- Date of Event
- August 18, 2014
- Report Date
- October 15, 2014
- Manufacturer
- THORATEC CORP.
- Product Code
- DSQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE SYSTEM CONTROLLER WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE REPORTED INCIDENT OF A DAMAGED BLACK POWER LEAD WAS CONFIRMED DURING VISUAL EXAMINATION. DURING ANALYSIS, THE DEVICE DID NOT PASS THE INSULATION HI-POT TESTING, WHICH INDICATED COMPROMISED UNDERLYING WIRES. MANIPULATION OF THE DAMAGED AREA RESULTED IN AN INTERRUPTION OF THE PUMP SPEED. VISUAL INSPECTION OF THE UNDERLYING WIRES REVEALED THE MECHANICAL INDUCED STRESS RESULTED IN THE SHIELDING TO PUNCTURE THE INSULATION, CAUSING A SHORT CIRCUIT CONDITION. THE LOG FILE RETRIEVED FROM THE SYSTEM CONTROLLER WAS REVIEWED AND CAPTURED MOTOR STOPPED EVENTS WITH ASSOCIATED LOW FLOW HAZARD (RED HEART) ALARMS. THE EVENTS WERE CAPTURED WHILE CONNECTED TO THE POWER MODULE CONSISTENT WITH THE ANALYSIS FINDINGS. A REVIEW OF THE DEVICE HISTORY RECORDS REVEALED THE SYSTEM CONTROLLER MET ALL APPLICABLE SPECIFICATIONS UPON PRODUCT RELEASE. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. THE MANUFACTURER IS CLOSING ITS FILE ON THIS EVENT.
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED BY THE VAD COORDINATOR THAT THE BEND RELIEF ON THE BLACK POWER LEAD OF THE SYSTEM CONTROLLER WAS DAMAGED. THE VAD COORDINATOR WAS UNABLE TO DOWNLOAD THE LOG FILE FROM THE SYSTEM CONTROLLER. THE SYSTEM CONTROLLER WAS EXCHANGED. DURING THE INVESTIGATION, DAMAGE TO THE BLACK POWER LEAD OF THE SYSTEM CONTROLLER WAS FOUND AND WHEN MANIPULATED PUMP STOPPAGE OCCURRED. PUMP STOPPAGE WAS ALSO NOTED IN THE LOG FILE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 732592 | HEARTMATE II SYSTEM CONTROLLER | DSQ: LVAD SYSTEM CONTROLLER | DSQ | THORATEC CORP. | 103696 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR |