FDA Adverse Event Summary report: N

BODYGUARD

MDR report key: 4253181 · Received September 24, 2014

Report

Report Number
4253181
Date Received
September 24, 2014
Date of Event
June 14, 2014
Report Date
September 11, 2014
Manufacturer
CME AMERICA, LLC
Product Code
FRN
Report Source
User Facility report
Reporter Location
LA, US

Narratives

Description of Event or Problem · 1

THE BRAUN IPX3 PCA PUMPS ARE: 1) SET UP TO BE USED IN ONE WAY ONLY, THAT IS USING A SMALL IV BAG; 2) THE PUMP CAN BE CONFIGURED WITH DIFFERENT PASSCODE REQUIREMENTS IN ORDER TO MAKE CHANGES (ALL, NONE, OR SOME); 3) THE LOCKABLE CASE HAS AN ACCESS WINDOW FOR THE KEYPAD: 4) THE PRIMING VOLUME IS NOT AUTOMATICALLY SUBTRACTED FROM THE STARTING VOLUME; AND 5) ONLY ONE TUBING SET WILL ALLOW THE USE OF PIGGY BACK INFUSION.WHEN "ALL" CODES ARE SET, THE NURSE MUST ENTER THE PASSCODE SEVEN (7) TIMES TO START AN INFUSION.THIS FACILITY HAD ELECTED TO USE SYRINGES, RATHER THAN BAGS, TO BETTER TRACK NARCOTIC USAGE. THE PLASTIC CASE HAD TO BE MODIFIED TO ACCOMMODATE THIS. IN ORDER TO REDUCE THE NUMBER PASSCODES REQUIRED A PLEXIGLASS COVER WAS INSTALLED OVER THE OPENING IN THE CASE.IF THE BEGINNING VOLUME IS NOT MANUALLY CORRECTED AFTER PRIMING, SUBSEQUENT INFUSED VOLUMES WILL NOT CORRESPOND WITH THE SYRINGE VOLUME.======================MANUFACTURER RESPONSE FOR PCA PUMP, CME COLOR VISION (PER SITE REPORTER).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
593176 BODYGUARD PUMP, INFUSION FRN CME AMERICA, LLC 575 COLORVISION *

Patients

Seq Age Sex Outcome Treatment
1 *