CHARGING SYSTEM
Report
- Report Number
- 1627487-2014-06343
- Event Type
- Injury
- Date Received
- November 11, 2014
- Date of Event
- October 16, 2014
- Report Date
- October 16, 2014
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULT: POCKET HEATING CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2014-06342. IT WAS REPORTED AN THE PATIENT EXPERIENCED AN UNCOMFORTABLE HEATING SENSATION A THIS IPG SITE WHILE CHARGING. A NEW LE CHARGER WAS SENT TO THE PATIENT TO ADDRESS THE ISSUE. ON 08/01/2012 ST JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 726516 | CHARGING SYSTEM | CHARGING SYSTEM | GZB | ST. JUDE MEDICAL - NEUROMODULATION | 3711 | 3556859 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Other | SCS LEAD: MODEL 3186 (2)| SCS ANCHOR: MODEL 1192 (2)| IMPLANT DATE:| IMPLANT DATE: |