FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4252495 · Received November 15, 2014

Report

Report Number
2939301-2014-30698
Event Type
Malfunction
Date Received
November 15, 2014
Report Date
November 5, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (12/23/2014). THE PATIENT¿S METER HAS BEEN RETURNED ON 11/18/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 12/11/2014 WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 2 - 1/28/2015 DEVICE EVALUATION. THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED ON 1/13/2015 AND FURTHER EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 1/21/2015 WITH THE FOLLOWING FINDINGS:THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE CONTROL SOLUTION HAS ALSO BEEN RETURNED TO LFS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING AN UNUSUAL ISSUE WITH THEIR ONETOUCH VERIO IQ METER. THE REPORTER STATED THAT WHEN THE METER BATTERY IS LOW THE METER GIVES INACCURATE RESULTS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.