10MM, 33CM RATCHETING HANDLE
Report
- Report Number
- 0002936485-2014-00930
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 23, 2014
- Report Date
- October 23, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING INSPECTION OF THE 10MM, 33CM RATCHETING HANDLE, DINGS AND SCRATCHES WERE NOTICED THROUGHOUT THE SHAFT. ALSO, CRACKS WERE SEEN ON BOTH SIDES OF THE DISTAL TIP OF THE SHAFT. IFU 10000401070 STATES, "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE-THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE/S COULD BE NORMAL WEAR, USER MISHANDLING OR IMPROPER STERILIZATION. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 736157 | 10MM, 33CM RATCHETING HANDLE | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 1042940D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |