FDA Adverse Event Malfunction Summary report: N

10MM, 33CM RATCHETING HANDLE

MDR report key: 4252357 · Received November 14, 2014

Report

Report Number
0002936485-2014-00931
Event Type
Malfunction
Date Received
November 14, 2014
Date of Event
October 23, 2014
Report Date
October 23, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION DURING VISUAL INSPECTION. THE PROBABLE ROOT CAUSES COULD BE NORMAL WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION DURING VISUAL INSPECTION. IFU 10000401070 "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE-THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSES COULD BE NORMAL WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736138 10MM, 33CM RATCHETING HANDLE ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 1045186D

Patients

Seq Age Sex Outcome Treatment
1