10MM, 33CM RATCHETING HANDLE
Report
- Report Number
- 0002936485-2014-00931
- Event Type
- Malfunction
- Date Received
- November 14, 2014
- Date of Event
- October 23, 2014
- Report Date
- October 23, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION DURING VISUAL INSPECTION. THE PROBABLE ROOT CAUSES COULD BE NORMAL WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DENTS AND SCRATCHES WERE SEEN ON THE INSULATION DURING VISUAL INSPECTION. IFU 10000401070 "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE-THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSES COULD BE NORMAL WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.